Sunum yükleniyor. Lütfen bekleyiniz

Sunum yükleniyor. Lütfen bekleyiniz

Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 1 Medical Instrumentation Application and Design, 4th Edition John Wiley and Sons Ltd, Feb 2009, Pages: 713.

Benzer bir sunumlar


... konulu sunumlar: "Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 1 Medical Instrumentation Application and Design, 4th Edition John Wiley and Sons Ltd, Feb 2009, Pages: 713."— Sunum transkripti:

1 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 1 Medical Instrumentation Application and Design, 4th Edition John Wiley and Sons Ltd, Feb 2009, Pages: 713 Basic Concepts of Medical Instrumentation Basic Sensors and Principles Amplifiers and Signal Processing The Origin of Biopotentials Biopotential Electrodes Biopotential Amplifiers Blood Pressure and Sound Measurement of Flow and Volume of Blood Measurements of the Respiratory System Chemical Biosensors Clinical Laboratory Instrumentation Medical Imaging Systems Therapeutic and Prosthetic Devices Electrical Safety

2 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 2

3 3

4 4

5 5

6 6

7 7

8 8

9 9

10 10

11 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 11

12 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 12 A 100A

13 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 13 Ödevler: Bioelektrik empedans tanımı, ölçmeleri, dijital terazide vücudun yağ, su değerlerinin belirlenmesi 1.Thermocouples Thermistor 3.Resistance Temperature Detector 4.Infrared Thermometers 5.Pressure Transducers 6.Load Cells 7.Strain Gage 8.Flowmeters 9.pH Measurement 10.Level Measurement 28 Ekim Kasım 2011

14 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 14

15 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 15

16 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 16

17 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 17

18 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 18

19 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 19 G H xy

20 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 20

21 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 21

22 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 22

23 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 23

24 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 24

25 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 25

26 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 26

27 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 27

28 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 28

29 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 29

30 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 30

31 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 31

32 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş  cardiac cycle Systole Diastole pressure volume flow 32

33 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 33 astr.gsu.edu/hbase/biology/ecg. html#c2 astr.gsu.edu/hbase/biology/hear telec.html#c1

34 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 34

35 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 35 Elektrik akımı ve geriliminin insan için zararlı, öldürücü olma değerlerini, frekans özelliklerini de göz önüne alarak anlatın.(%20) Duyarlık elemanlarımızın cevaplarının zaman göre çıkışlarının: (giriş çıkış arasındaki gecikme var/yok, gecikmenin türü) zero-order (%5) firs-order (%5) second-order (%5) olma özelliklerini anlatın. EKG işaretlerinin algılanmasını, elde edilen işaretin genlik ve frekans özelliklerini, şeklini ve anlamlarını anlatın. (%20) Kan basıncını (mekanik-analog) ölçme yöntemini (ölçülen değerlerin anlamlarını belirterek) anlatın. (%15) Bioelektrik empedans tanımı, ölçmeleri, dijital terazide vücudun yağ, su değerlerinin belirlenmesi anlatın. (%20) Sadece üst tarafından erişilebilen 10 m derinlikte bir havuzunun içindeki suyun seviyesiini en güvenli şekilde ölçmek, ve display etmek için gerekli farklı iki sistemi tasarlayı, çizin, anlatın. Öneri- 1(%5), Öneri-2(%5),

36 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 36

37 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 37

38 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 38

39 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 39 Miniature Wireless Pulse Oximeter esearch:wireless_biosignal_telemetry china.com/offer/abHEQvnYztpJ/Sell- CMS-P-PC-Based-Pulse-Oximeter- with-Free-SW.html CMS-P PC Based Pulse Oximeter

40 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 40

41 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 41 The probe contains two high intensity, monochromatic, light- emitting diodes, one emitting red light (660 nm) and the second infrared (940 nm) on one side and a photodetector on the other to measure the amount of light transmitted through the finger. Pulse oximetry

42 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 42 based-heart-rate-meter/

43 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 43 CD4046-CD4511.aspx

44 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 44

45 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 45

46 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 46

47 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 47

48 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 48

49 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 49

50 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 50

51 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 51

52 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 52

53 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 53

54 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 54 w/lung-vol.html

55 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 55

56 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 56

57 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 57

58 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 58

59 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 59

60 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 60 Ödevler: Bioelektrik empedans tanımı, ölçmeleri, dijital terazide vücudun yağ, su değerlerinin belirlenmesi 1.Thermocouples Thermistor 3.Resistance Temperature Detector 4.Infrared Thermometers 5.Pressure Transducers 6.Load Cells 7.Strain Gage 8.Flowmeters 9.pH Measurement 10.Level Measurement

61 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 61 this multimeter may also be known as a volt-ohm-milliammeter

62 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 62

63 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 63

64 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 64 (1) Each units are: t seconds, R ohms, C farad and epsilon is a Napier's number (approx. 2.72). When VC reaches VC1, the time t1 can be expressed in following formure. (2) This means that the t 1 is proportional to C. Thus the capacitance can be calcurated from charge time and any other fixed parameters. Digital Capacitance Meter

65 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 65 Displacement, Location, or Position Sensors

66 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 66 Displacement, Location, or Position Sensors

67 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 67

68 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 68 ANALOG FREQUENCY MEASUREMENTS

69 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 69

70 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 70 This audio frequency meter uses 555 IC as a monostable multivibrator (one-shoot trigger). A monostable multivibrator can act as a frequency-to-voltage converter because it produce a fixed pulse width, with the repetition rate/density is proportional to the triggering input frequency. Here is the circuit’s schematic diagram: For resistor R1, because it set the measurement range, it’s better to use a rotary switch to select different values for different ranges. For the ampere meter, you can use both analog or digital ampere meter. A cheap dual-slope ADC digital meter is suitable because its averaging characteristic, but a fast digital multimeter can also be used although it may show some uncertainty because of their fast sampling.

71 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 71 This 1-kHz linear-scale analog frequency meter circuit uses the 555 as a pulse counter

72 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 72

73 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 73 DEFIBRILLATORS

74 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 74 For this example, the discharge is underdamped (biphasic, also referred to as a Gurvich waveform) when the patient resistance is less than about 56Ω because Rpatient+Rinductor=56Ω+10Ω=66Ω

75 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 75 Figure 8.32 Schematic diagram of a damped sinusoidal waveform defibrillator capable of delivering energies of up to 320 J into a 50-Ω patient load through a 5-ms Edmark (monophasic) waveform. Charge pushbutton SW2 energizes high-voltage transformer T1. C1 is charged through the high-voltage rectifier network D1–D4 and R1. Meter M1 is calibrated to yield an estimate of energy (in joules) delivered to the patient, assuming a load impedance of 50Ω. Defibrillation energy is delivered to the patient by simultaneously pressing on pushbuttons SW3 and SW4, which energize relay K1, which is used to transfer the defibrillation charge from capacitor C1 to the patient via pulse shaping inductor L1. R4 and R5 discharge C1 if the defibrillation buttons are depressed without a suitable load across the paddle electrodes or the

76 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 76

77 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 77 Energy limits A.C. defibrillator 0.25 sec × 230 volts × (I Amps) = 450 watts-sec. Required required I = 7.5Amps. But preferable 0.01 sec. Pulses. Then I = 450/(230 × 0.01) = 175 A. This is too much to be supplied by household mains.

78 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 78 DC Defibrillation (Fig b) 1. Safety housings for electrodes–capacitor discharges only when the electrodes are making firm contact with the heart or chest wall. 2. Two set of electrodes - not interchangeable sockets 1. Internal 50–72 J (5–3 kV) 3. Meter indicates Joules 1. External 400J (7 kV). 4. Charging time constant of 4 seconds M Ω × 16 μF 2. (charging resistor) Takes about 16 secs. to charge to 4 kV.

79 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 79 CAPACITIVE-DISCHARGE DC DEFIBRILLATORS A short high-amplitude defibrillation pulse can be obtained by using the capacitive-discharge circuit shown in Figure In this case, a half- wave rectifier driven by a step-up transformer is used to charge the capacitor C. A good rule of thumb is to keep charging time under 10 s. MEDICAL INSTRUMENTATION Application and Design FOURTH EDITION John G. Webster, Editor

80 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 80

81 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 81 When external electrodes are used, energies as high as 400 J may be required. The energy stored in the capacitor is given by the well- known equation

82 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 82

83 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 83 ts/type/emergency/tec7700.html Fast charging — less than 3 seconds to 150J and less than 5 seconds to 270J

84 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 84

85 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 85

86 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 86

87 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 87 Principles of Bioelectrical Impedance Analysis

88 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 88

89 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 89 A method which involves the measurement of bioelectrical resistive impedance (R) for the estimation of human body composition is described. This method is based upon the principle that the electrical conductivity of the fat-free tissue mass (FFM) is far greater than that of fat. Using an electrical impedance plethysmograph with a four electrode arrangement that introduces a painless signal (800  A at 50 kHz) into the body. FFM was assessed by hydrodensitometry and ranged from kg. Total body water (TBW) determined by D2O dilution and total body potassium (TBK) from whole body counting were 50.6 ± 10.3 L and ± 38.1 g, respectively..

90 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 90

91 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 91

92 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 92 Tepe idrar akım hızı erkeklerde 20 ml/sn, kadınlarda ise 25 ml/sn üstünde normal oalrak kabul edilir Uroflowmetry The SEDIA F1 Flowmeter provides you with a convenient, cost effective and easy to use solution for portable uroflowmetry. Communication between the flow meter and your computer is via Bluetooth connection. The SEDIA software is easily installed on your Windows PC, allowing you to print reports from your own printer. urodynamics.htm

93 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 93 /flow814-graph-zoom.html

94 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 94 Turbine Flowmeters for Liquid Measurement tutorials/flowmetering/types-of-steam-flowmeter.asp

95 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 95 Peak-Flow Metre (Nefes ölçüm testi)

96 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 96 Peak-Flow Metre (Nefes ölçüm testi)

97 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 97 Breath Alcohol Tester flashlight-keychain-ce277777

98 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 98 Alcohol Tester Sensor MQ303A is semiconductor sensor is for Alcohol detection, Model No.MQ303A Sensor TypeSemiconductor Standard EncapsulationMetal Detection GasAlcohol Concentration ppm Alcohol Standard Circuit Conditions Heater VoltageV H0.9V ± 0.1V AC or DC Loop VoltageVc≤6V DC Load ResistanceR LAdjustable Heater Resistance R H4.5W ± 0.5 W(Room Tem.) Heater CurrentIH120±20mA Heater PowerPH≤ 140 mW Character Sensor Consumption PS≤10 mW Sensing Resistance Rs4KΩ-400KΩ(in air) SensitivitySRs(in air)/Rs(125ppm Alcohol)≥3 Slopeα0.50 ± 0.15(R300ppm/R100ppm Alcohol) Condition Tem. Humidity20°C±2°C;65%±5%RH Standard test circuit Vc:3.0 V±0.1 V DC; VH: 0.9 V±0.1 V DC Preheat timeOver 48 hours heets/ITEAD/MQ303A.pdf

99 Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş

100

101 101 Digital Blood Pressure Meter rs/doc/app_note/AN1571.pdf


"Prof.Dr.Sezgin ALSAN Biyomedikal Müh. Giriş 1 Medical Instrumentation Application and Design, 4th Edition John Wiley and Sons Ltd, Feb 2009, Pages: 713." indir ppt

Benzer bir sunumlar


Google Reklamları